Calcium Hradness

F R E E C H L O R I N E

F R E E C H L O R I N E

F R E E C H L O R I N E

F R E E C H L O R I N E

F R E E C H L O R I N E

SPECIFICATIONS

Range

0.00 to 5.00 mg/L

Resolution

0.01 mg/L from 0.00 to 2.50 mg/L;
0.10 mg/L above 2.50 mg/L

Accuracy

±0.03 mg/L ±3% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the USEPA method 330.5 and

Standard Methods for the Examination of

Water and Wastewater, 20

th

edition, 4500-Cl G. The reaction between free chlorine and

the DPD reagent causes a pink tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93701-0

DPD powder Reagent

1 packet

REAGENT SETS

HI 93701-01 Reagents for 100 tests
HI 93701-03 Reagents for 300 tests
For other accessories see page 46.

MEASUREMENT PROCEDURE

• Select the

Free Chlorine method using the procedure described in the

Method Selection section (see page 15).

• Fill the cuvette with 10 mL of unreacted sample (up to the mark)

and replace the cap.

• Place the cuvette into the holder and close the lid.

Free Chlorine

FREE CHLORINE

Note

: This test will detect any calcium contamination in the beaker, measuring syringes or sample cells. To

test cleanliness, repeat the test multiple times until you obtain consistent results.

Note: For better accuracy wash glassware with HCl 6N.

INTERFERENCES

Interference may be caused by excessive amounts of heavy metals.

• Press the

Unit key to change the current measurement unit. The results can be converted to French

degrees (°f), German degrees (°dH) and English degrees (°E).

• Press ▲ or ▼ to return to the measurement screen.

10 mL