• Immerse the temperature probe in the second vessel as near as possible to the reference

thermometer. Allow a few seconds for the probe to stabilize.

• Use the ARROW keys to set the calibration point value to that of the hot water, measured by

the reference thermometer.

• The

(unstable measurement) symbol will be shown on the display until the reading

becomes stable.

• When the reading is stable and close to the selected calibration point, the (unstable

measurement) symbol will disappear and the Confirm key will become active.

• Press Confirm to confirm the calibration point. The instrument will return to the pH meter/

titrator main screen.

Notes:• If the reading is not close to the selected calibration point, the “Wrong” message will

be displayed. Change the temperature probe and restart the calibration.

• If the temperature probe is disconnected or the measured temperature is out of the -

20 to120 °C range the instrument will display “----”. The calibration point value can
be changed using the ARROW keys.