Hanna Instruments HI 84442 User Manual
Page 17

17
• The
(unstable measurement) symbol will be shown on the display until the reading
becomes stable.
• When the reading is stable and close to the selected buffer, the (unstable measurement)
symbol will disappear and the Confirm key will become active.
• Press Confirm to confirm the calibration.
• Press ESC to exit calibration.
Notes:• To clear a previous calibration and to return to the default value, press Clear at any
time after entering calibration mode. The “Calibration cleared” message will be
shown for a few seconds on the display. If Clear is invoked during the first calibration
point the instrument returns to the measurement mode.
• The Clear key is displayed only if a previous calibration exists.
Point 2 calibration
• The calibrated value will be shown on the display and the second expected buffer value will
be displayed.
• Remove the electrode holder with electrodes from the top of the beaker.
• Place the second buffer into beaker and place in beaker holder. Rinse the electrodes in a
beaker containing the second buffer rinsing solution.
• Place the electrode holder (with electrodes) on the top of the beaker, lock cap by turning.
• If necessary press the ARROW keys in order to select a different buffer value.
• The (unstable measurement) symbol will be shown on the display until the reading becomes
stable.
• When the reading is stable and close to the selected buffer, the (unstable measurement)
symbol will disappear and the Confirm key will become active.
• Press Confirm to confirm the calibration.