Verify: The electrode has been calibrated in 4.01 pH buffer.
• Sample preparation: Add 1 mL of HI 84442-55 Pump

Calibration Solution to a clean beaker.
Fill the beaker up to the 75 mL mark with the distilled or

deionized water. Place the stir bar into the beaker and

then place the beaker into the appropriate place on the
instrument top.

• Place the electrode holder on the top of the beaker and

secure it by turning clockwise.

• Before immerse the electrodes in the beaker:

- Carefully rinse the pH electrode and temperature

probe with deionized water.

- Dry electrodes using soft paper.
- Rinse again with the deionized water.

• Immerse the calibrated pH and the temperature electrodes

approximatively 2 cm (0.8”) into the sample to be tested

paying attention not to touch the stir bar.

• Insert the dosing tip in the appropriate guide tube holder

place taking care not to empty the tip.

Note: The chemical reagents may be hazardous if improperly

handled. Read the Material Safety Data Sheets (MSDS)
before performing the test.

•From the titrator main screen press CAL.

After 5 minutes purging interval has elapsed, the “Completed” message is displayed. Another
purge period can be initiated by pressing Restart or press ESC to return to main titrator screen.

PUMP CALIBRATION PROCEDURE

The pump calibration must be performed each time the titrant bottle or the pH electrode is

changed. It is recommended the pump calibration to be performed before each set of titrations.