Operational Guide

pH Buff er
Temperature
Dependence

Note: To clear a previous calibration
and to return to the default value,
press CLEAR at any time after entering
calibration mode. If CLEAR is invoked
during the fi rst calibration, point the
instrument returns to the measurement
mode. The CLEAR key is displayed only if
a previous calibration exists.

Temperature has an eff ect on pH. The calibration buff er solutions are
aff ected by temperature changes to a lesser degree than normal solutions.
During calibration, the instrument will automatically calibrate to the pH value
corresponding to the measured or set temperature.

During calibration the instrument will display the pH buff er value at 25 ºC.

Temp

pH Buff ers

ºC

ºF

4.01

6.00

8.30

10.01

0

32

4.01

6.12

8.48

10.32

5

41

4.00

6.09

8.44

10.24

10

50

4.00

6.06

8.41

10.18

15

59

4.00

6.04

8.37

10.12

20

68

4.00

6.02

8.33

10.06

25

77

4.01

6.00

8.30

10.01

30

86

4.02

5.99

8.27

9.96

35

95

4.03

5.98

8.24

9.92

40

104

4.04

5.97

8.21

9.88

45

113

4.05

5.97

8.18

9.85

Electrode
Calibration
Procedure