Pump calibration messages – Hanna Instruments HI 84529-01 User Manual

Page 21

background image

www.

hannainst

.com

21

Setup/Installation

Pump
Calibration
Messages

This error message appears during pump
calibration when the endpoint cannot
be reached and the maximum amount
of titrant is exceeded. Check standard,
electrode and/or dosing system and try
again

The

calibration

was

outside

the

acceptable limits. Prepare a new standard
and try again.

This error message appears when the
pH reading exceeds the acceptable input
limits (pH <-2.00 or pH> 16.00).

This screen appears when the stirrer is
not working properly. Check the stir bar
and beaker content. Press RESTART to try
again.

This error message appears when the
pump is not working properly. Check the
tubing, valve and syringe. Press RESTART
to try again.