Method ID: HI9301EN

Moisture Determination in Solvent

for external dissolution or extraction

Titrator Application Methods: General Applications – r. 1.0

Description:
Method for the determination of moisture in
extraction/dissolution solvent. The results are
expressed in mg/g and should be less than 1.00
mg/g. Extractions/dissolutions of substances with
low water contents may require very dry solvents
(<0.100 mg/g).

Electrode:

 HI 76330 Double Platinum Pin Electrode
 HI 900517 Generator with Diaphragm

-or-

 HI 900518 Generator without Diaphragm

Reagents:
 General Purpose Coulometric Karl Fischer

Reagent

Other Accessories:
 A clean, dry 1-mL syringe
 A clean, dry 22-gauge, 6” non-coring septum

penetration needle

 GL 45-Thread Reagent Bottle

Procedure:
 Setup titrator according to the instruction manual.

 Press “Select Method” from the main screen. Use

the arrow keys to highlight ‘HI9301EN Moisture in
Solvent’ and press “Select”.

 Connect the reagent bottle top assembly to the

bottle of reagent according to the manual.

 Prepare the titration vessel according to the

manual. Dispense enough reagent from the
reagent bottle to fill the vessel to the “min” line
(about 100 mL).

 Press “Start/Stop” to pre-titrate the reagent and

the titration vessel moisture. Allow the
background drift rate to stabilize before running a
titration.

 Prepare an extraction/dissolution vessel with

solvent and stir (see an applicable standard
method for proper solvent amount and stirring
time).

 Stop stirring the solvent in the extraction/

dissolution bottle.

 Fill the syringe and needle with the

extraction/dissolution solvent.

 Weigh the syringe, needle and solvent.

 Press “Start Analysis”. You will be prompted to

enter the sample size.

 Dispense 0.8000 g to 1.0000 g of solvent into the

titration vessel through the septum using the
needle. Pay attention not to get any sample on
the electrode or beaker walls. If necessary swirl
the titration vessel by hand.

 Clear the needle of residual sample by in taking a

small volume of air from the titration vessel. If a
“hanging drop” of solvent is seen on the end of the

needle, dip the end of the needle briefly in the
solvent.

 Remove the needle from the titration vessel and

weigh the syringe again in order to determine the
added sample mass (by difference of the two
measurements.)

 Use the numeric keypad to enter the exact weight

and press “Enter” to start the analysis.

 At the end of the titration the “Sample Analysis

Result” screen is displayed. The results are
expressed in mg/g of water.

Method Parameters:

Name:

Moisture in Solvent

Method Revision:

1.0

Type:

KF Coulometric

Pre-Analysis Stir Time:

5 Sec

Stirring Speed:

900 RPM

Stirbar Type:

Medium

Drift Entry:

Automatic

Reagent:

General Purpose

Sample Parameters:

Sample Determ.:

Normal

Sample Name:

Solvent

Sample Type:

Mass

Sample Size:

1.0000 g

Control Parameters:

Titration Speed:

Auto

Standby Mode:

Enabled

Standby Duration:

12:00 [hh:mm]

Imposed Current:

2uA

End Point Value:

100.0 mV

Generator Current Mode:

Auto

Signal Averaging:

2 Readings

Termination Parameters:

Maximum Duration:

1200 sec

Maximum Titrated Water:

10.000 mg

Term. Criterion:

Relative Drift

Relative Drift:

3.0 μg/min

Result Unit:

ppm

Calculations:

Water titrated:

H

2

O (μg)

Final results units:

ppm

Sample mass:

1.0000 g

ppm =

water (μg

)

1.0000

g

Results:

Titration Report

Method Name:

Moisture in Solvent

Time & Date:

12:00 Jan 01, 2013

Sample Size:

1.0322 g

Drift Value:

0.9ug/min

Titrated Water:

208.10 ug

Result:

200.1ppm

Titration Duration:

1:45 [mm:ss]

Generator Electrode Type:

HI 900518

Titration went to Completion
Operator Name:

Analyst Signature: _______________________