beautypg.com

Methods optimization – Hanna Instruments HI 904 User Manual

Page 115

background image

9-3

METHODS OPTIMIZATION

9

OPTIMIZATION

9.1

Titration Settings

This section provides the descriptions of critical titration parameters necessary for an analyst
to modify a standard method or develop a titration method from scratch.

HI 904 methods can be modified and customized based on the requirements of the sample,
sample matrix and the Karl Fischer reagent formulation. The user changeable settings are
separated into two categories: Control Parameters, which set critical functions that determine
the course of a titration and set the way in which titrations are terminated, and Method
Options, which control lesser features not directly affecting measurements and primarily
allow advanced users to shorten titration times.

9.1.1

Control Parameters

9.1.1.1 Endpoint Potential and Polarization Current

The HI 904 uses the polarized electrode system known as bivoltametric indication. The
titrator monitors the voltage required to maintain a pulsed polarization current (Ipol) between

the pins of a dual platinum-pin Karl Fischer electrode during the course of a titration.
During a titration, no excess iodine is present. In order to maintain the set polarization
current the HI 904 must apply a relatively large voltage across the pins of the electrode.
At the endpoint of the titration, the amount of iodine added is equal to the amount of water
from the sample. When an excess of iodine has been generated the excess iodine is easily
reduced, and the resulting iodide is easily oxidized in electrode reactions at the cathode and
anode respectively. The ease of these reactions make maintaining the polarization current
possible at a much lower electrode potential.
In theory, a large shift in the electrode potential indicates the endpoint. In practice, a titration
endpoint is reached when the electrode potential drops below a value defined by the user
and the chosen termination criteria is met.
The choice of endpoint potential should be based, foremost, on the polarization current and,
to a lesser extent, on the composition of the Karl Fischer reagent and the sample matrix. If
the polarization current is changed, the endpoint potential must also be changed. In addition,
there are pitfalls to be avoided when choosing an endpoint potential. Selecting endpoints
which are both ‘too high’ or ‘too low’ will result in long titration times and poor reproducibility.
Endpoints which are ‘too high’ are those which result in endpoints that either precede or
coincide with equivalence point such that the concentration of excess iodine is not reliably
detected. Endpoint potentials are considered ‘too low’ when they correspond to a large
excess of iodine in the titration cell.