QUICK START GUIDE
BEFORE PERFORMING THE FIRST TITRATION
Setup the Titrator
• Make sure that all of the titrator assemblies are properly installed (see Instruction
Manual,
Setup section).
• Make sure that the vessel system is properly sealed against atmospheric moisture
(the fittings and tubes are correctly mounted).
• The desiccant had been properly dried.
Obtaining the Reagents
• The reagents have to be suitable to the analysis requirements (see Instruction Manual,
Appendix 2 for list of preferred reagents).
HOW TO PERFORM THE FIRST TITRATION
Method Selection
For this analysis we will use the HI 9301EN Moisture in Solvent
.
To select this method:
• Press
from the
Idle screen. Use the
and
keys to highlight the
HI 9301EN Moisture in Solvent method.
• Press
.
After accomplishing these operations, the method’s name will be displayed on the top line of
the
Idle screen.
Setup Titration Report
Users can select the information that is stored for each titration that is performed.
To obtain proper information at the end of the titration, perform the following operations:
• From the main screen, press
and the
Data Parameters screen will be
displayed.
• Highlight the
Setup Titration Report option and press
.
• Mark the fields to be included with the “*” symbol using the
and
keys and
press
to toggle the selection.
• Press
and then press
to return to the main screen.