29

Press

GLP

to enter or exit GLP data consulting.

Several functions are available when in GLP menu.

Press

READ to scroll the following GLP data:

• The last calibration date, in YYYY.MM.DD

format. If no calibration was performed, the
factory calibration message, “

F.CAL”, will be

displayed on the LCD.

• The time of the last calibration in hh:mm

format.

• First calibration point: 0.00 NTU if skipped or

the actual read value (e.g. 0.01 NTU).

• Second calibration point.

GOOD LABORATORY PRACTICE (GLP)

• The instrument asks for confirmation and if

LOG/CFM is pressed, all records are deleted.
To abort the delete function, press

READ

instead of

LOG/CFM.

• After all records are deleted, dashes are

displayed for one second and the instrument
returns to idle mode.

The GLP feature allows the user to view last calibration data. Also, the user calibration can be
deleted.