beautypg.com

Hanna Instruments HI 98703 User Manual

Page 15

background image

In order to match multiple cuvettes follow the next
steps:
• Fill some cuvettes with high quality water

(<0.1 NTU) up to the line.

• Clean and oil the cuvettes as described before.

• Turn the instrument ON.

• Insert the first cuvette into the instrument and

press

READ . Record the reading.

• Record the position of the cuvette and the

displayed reading.

• Mark this position on the thicker white band on

the top of the cuvette with a water resistant
pencil.

• Insert the second cuvette into the instrument

and take a reading.

MATCHING MULTIPLE CUVETTES

Precise measurements require the use of a single cuvette. If it is not possible, the cuvette
selection and matching must be performed before taking measurements.

15