• At the end of the measurement, the third

calibration point (100 NTU) is displayed on
the primary LCD, “

CAL P.3” on the secondary

LCD, and “

READ” tag will blink.

• At this moment it is possible to exit calibration

by pressing

CAL. The instrument will memorize

the two-point calibration data and will return
to measurement mode.

THREE-POINT CALIBRATION

To perform a three-point calibration, continue the
procedure with the following steps:
• Remove the second standard cuvette.
• Place the 100 NTU standard cuvette (or the

prepared formazin standard) into the holder,
with the cuvette mark aligned with the sign on
the instrument.

• Close the lid and press

READ . The display

will show blinking dashes and the icons for
cuvette, detectors and lamp will appear during
measurement.

• At the end of the measurement, the fourth

calibration point (750 NTU) is displayed on
the primary LCD and “

CAL P.4” on the

secondary LCD.

• At this moment it is possible to exit calibration

by pressing

CAL. The instrument will memorize

the three-point calibration data and will return
to measurement mode.

FOUR-POINT CALIBRATION

To perform a four-point calibration, continue the
procedure with the following steps:
• Remove the third standard cuvette.
• Place the 750 NTU standard cuvette (or the

prepared formazin standard) into the holder,
with the cuvette mark aligned with the sign on
the instrument.