In order to index a cuvette follow the next steps:
• Fill the cuvette with high quality water (<0.1

NTU) up to the line.

• Clean and oil the cuvette as described before.

• Turn the instrument ON.

• Insert the cuvette into the instrument and press

READ . Record the reading.

• Open the instrument lid, slightly rotate the

cuvette and take a new reading.

• Repeat the last step until you read the lowest

NTU value. Alternatively, keep

READ pressed

and, after the first value is displayed, open the
lid and start rotating the cuvette until the
lowest NTU value is displayed.

• Mark this position on the thicker white band on

the top of the cuvette with a water resistant
pencil.

• Always use this position to align it with the

sign on the instrument top.