4

1.0 Mol L

-1

Na

+

0.1 Mol L

-1

Na

+

Sodium Ion Correction for the Glass at 20-25 °C

Concentration

pH

Error

13.00

13.50

14.00

12.50

13.00

13.50

14.00

0.10

0.14

0.20

0.10

0.18

0.29

0.40

Since the resistance of the pH electrode is in the range of 50-200 Mohm

depending on the composition of the glass, the current across the membrane

is in the pico Ampere range. Large currents can disturb the calibration of the

electrode for many hours.

For these reasons high humidity environments, short circuits and static dis-

charges are detrimental to a stable pH reading.

The pH electrode’s life also depends on the temperature. If constantly used at

high temperatures, the electrode life is drastically reduced.

Typical Electrode Life

Ambient Temperature

1- 3 years

90 °C

Less than 4 months

120 °C

Less than 1 month

Alkaline Error
High concentrations of sodium ions interfere with readings in alkaline solu-

tions. The pH at which the interference starts to be significant depends upon

the composition of the glass. This interference is called alkaline error and causes

the pH to be underestimated. Hanna’s glass formulations have the indicated

characteristics.