CAL

C

containing ice and water. Wait a few minutes for the probe to stabilize.

• Press the CAL and LOG keys simultaneously and then switch the instru-

ment on. After a few seconds, the instrument enters the temperature

calibration mode. The “CAL” tag will come on. The primary LCD will

display the measured temperature or the ”- - - -” message if the

measured temperature is out of range. The secondary LCD will display 0.0 °C
(First calibration point). The “ ”tag will blink until the reading has

stabilized.

• Use the ARROW keys to set the reading on the secondary LCD to that of

ice and water measured by ChecktempC (for example, -0.1 °C).

• When the reading is stable and close to the selected calibration point, the

“CFM” tag will blink.

• Press the CFM key to confirm the calibration or the CAL key to leave the

calibration mode.

• The secondary LCD will display 50.0 °C (Second calibration point).
• Immerse the temperature probe and the ChecktempC into the second

vessel.

• Allow a few minutes for the probe to stabilize.
• The “ ” tag will blink until the reading has stabilized.
• Use the ARROW keys to set the reading on the secondary LCD to that of

the hot water bath.

• When the reading is stable and close to the selected calibration point, the

“CFM” tag will blink.

• Press the CFM key to confirm the calibration or the CAL key to leave the

calibration mode.

Note: If the measured value is not close enough to the displayed value on

secondary LCD, the “WRONG “ tag will blink. In this case, check if the

value set on secondary LCD and the temperature measured with

ChecktempC are close. Change the temperature probe and restart

calibration if necessary.

CAL

CFM

C

-0.1°C