Warning messages – Hanna Instruments HI 221 User Manual

Page 13

background image

13

2

CAL

CLEAR CAL if new electrode

ATC

BUFFER 1

pH

ATC

CAL

CFM

BUFFER 1

pH

ATC

ELECTRODE

CLEAN

ELECTRODE
This message appears if the cleaning procedure performed as a result of the

above two messages is found by the instrument to be unsuccessful. Replace the

electrode.

WRONG BUFFER TEMPERATURE
This message appears if the temperature of the buffer is outside the defined
buffer temperature range.

WARNING MESSAGES

During calibration, the Calibration Check feature analyzes the electrode calibration

history and warns the user when problems have been detected. It is possible to

over ride the warning messages and confirm the calibration but it is not

recommended.
CLEAR CAL IF NEW ELECTRODE
This warning is displayed any time the new calibration parameters are better

than the previous parameters. You can clear the calibration history by press-

ing the CLR key, or continue by pressing the CAL key.

CLEAN ELECTRODE
This warning appears during Calibration Check for the second calibration

buffer when the instrument has detected a small variation of offset or both

offset and slope parameters. This variation may result from dirt on the

electrode. Refer to the electrode cleaning procedure. This ensures the removal

of film, dirt or deposits on the glass bulb and reference junction.

CLEAN ELECTRODE alternating with CHECK BUFFER
This warning appears during Calibration Check in the first calibration buffer as

CAL

WRONG

CLEAR CAL if new electrode

ATC

pH

BUFFER