Error messages, Enhanced calibration messages – Hanna Instruments HI 221 User Manual

Page 12

background image

12

CAL

BUFFER 1

pH

ELECTRODE

CLEAN

ATC

ATC

The stored calibration history to used issue error and warning messages during

calibration to help ensure the highest accuracy.
As electrode aging is normally a slow process, substantial changes from

previous calibrations are likely due to a temporary problem with the electrode

or buffers. Calibrating under these conditions will give measurement errors.

ERROR MESSAGES

Error messages appear if one or all of the calibration parameters are out of

accepted windows. When these messages are displayed calibration cannot be

confirmed.
WRONG BUFFER
This message appears when the difference between the pH reading and the

value of the selected buffer is too big. If this error message is displayed, check

if you have selected the proper calibration buffer.

CLEAN ELECTRODE
This error message indicates a bad electrode condition (offset out of accepted

window, or slope under the accepted lower limit).
Clean the electrode according to the Cleaning Procedure on page 32 to

improve its condition and repeat the calibration. This ensures the removal of

film, dirt or deposits on the glass bulb and reference junction.

CHECK ELECTRODE alternating with CHECK BUFFER
This error message appears when electrode slope exceeds the highest accepted

slope limit. You should check your electrode and use fresh buffer.