beautypg.com

Hanna Instruments HI 221 User Manual

Page 15

background image

15

4

GLP is a set of functions that allows the storage and retrieval of data regarding

the maintenance and status of the electrode.
All data regarding the last calibration (one or two points) is stored for the user

to review when necessary. This data includes the following: calibration time

stamp, offset (in mV), slope (in mV/pH), electrode condition and response

gauges, calibration buffers and the amount of time until a new calibration is

required.
CALIBRATION ALARM TIME-OUT
HI 221 and HI 223 allow the user to set the number of days before the next

required pH calibration. This value can be set from 1 to 7 days. The default

value is OFF (disabled).
The instrument checks if the time-out time has expired. If the time has

elapsed, the CAL DUE will blink as a reminder.
Note: If the instrument was not calibrated, or if the calibration history was

deleted, the CAL DUE message will be displayed even if this feature is

disabled in the SETUP menu.
If the instrument was calibrated using an electrode with pin and the

electrode is changed with an electrode without pin or vice-versa CAL

DUE will blink. This feature helps ensure use of a calibrated instru-

ment.

LAST CALIBRATION DATA
Last calibration data is stored automatically after a successful calibration.
To view the pH calibration data, press the GLP key when the instrument is in

pH measuring mode.
The instrument will display the time of the last calibration.
Press the ARROW keys to view the following logged calibration parameters

(pressing the / ºC key):
_ The time (hh:mm).

GOOD LABORATORY PRACTICE (GLP)

Also, when the instrument cannot evaluate the electrode response or pH 1.68 or
pH 12.45 buffer were used as calibration buffer, the response gauge will be empty.
If the electrode is in a very poor condition the first condition segment will blink.
If electrode response is very slow the first response segment will blink.

This manual is related to the following products: