beautypg.com

Hanna Instruments HI 221 User Manual

Page 10

background image

10

ATC

PROCEDURE
The user has a choice of 7 memorized buffers: pH 1.68, 4.01, 6.86, 7.01,

9.18, 10.01 and 12.45.
It is always recommended to perform a two-point calibration, however the

instrument also allows a one-point calibration, as described on page 11.

TWO-POINT CALIBRATION
For most applications it is recommended that pH 7.01 or 6.86 buffers be used

as the first calibration point and pH 4.01 (for acidic samples) or pH 9.18/

10.01 (for alkaline samples) as the second calibration point.
Note: The pH 12.45 buffer is not for general measurement; use only if the

sample is very alkaline to avoid sodium error.

• Immerse the pH electrode and the temperature probe approximately

4 cm (1½”) into the buffer solution of your choice (pH 1.68, 4.01,

6.86, 7.01, 9.18, 10.01 or 12.45) and stir gently. The temperature

probe should be close to the pH electrode.

• Press the CAL key. “CAL” and “pH” tags will be on, and the “CLEAR

CAL if new electrode” tag will blink.

• Press the CLR key if you are using a new electrode or want to clear the

calibration history. The instrument will display the “donE” message for

a few seconds.

It is very important to clear the calibration history when a new electrode is

used because all error and warning messages that appear during calibration

depend on the calibration history.

• Press the CAL key, or wait a few seconds to continue.

Note: The above behavior happens only if calibration history is not empty.

• The instrument will display the measured pH on the primary LCD, and

the most common buffer (“7.01”) on the secondary LCD along with

“CAL”, “pH” and “BUFFER 1” tags.

• The “ ” tag will blink until the reading has stabilized.

CLEAR CAL if new electrode

pH

CAL

This manual is related to the following products: