REQUIRED REAGENTS
Code
Parameter
Description
Quantity/test
HI 93710-0
pH
Phenol red
5 drops
HI 93701-0
Free Chlorine
DPD Powder Reagent
1 packet
HI 93711-0
Total Chlorine
DPD Powder Reagent
1 packet
HI 93722-0
Cyanuric Acid
Powder Reagent
1 packet
HI 93718-0
Iodine
DPD Powder Reagent
1 packet
HI 93716-0
Bromine
DPD Powder Reagent
1 packet
HI 93746-0
Iron LR
Powder Reagent
2 packet
Light Source
Tungsten lamp
Light Detector
Silicon Photocell with narrow band interference filter @ 525 nm
Methods
For pH: Phenol red method. The reaction with reagents causes a red
tint in the sample.
For Chlorine: Adaptation of the USEPA method and Standard Method
4500-Cl G. The reaction with reagents causes a pink tint in the
sample.
For Cyanuric Acid: Adaptation of the turbidimetric method. The reaction
between cyanuric acid and the reagent causes a white suspension in
the sample.
For Iodine: Adaptation of the EPA, DPD method. The reaction between
iodine and the reagent causes a pink tint in the sample.
For Bromine: Adaptation of the EPA, DPD method. The reaction
between bromine and the reagent causes a pink tint in the sample.
For Iron LR: Adaptation of the TPTZ method. The reaction between
iron and the reagent causes a violet tint in the sample.
Environment
0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type
1 x 9 volt
Auto-Shut off
After 10' of non-use in
measurement mode
;
after 1 hour of non-use in
calibration mode
;
with last reading reminder.
Dimensions
192 x 104 x 69 mm (7.6 x 4.1 x 2.7")
Weight
360 g (12.7 oz.).