Functional description – Hanna Instruments HI 96101 User Manual
Page 10
10
KEYPAD DESCRIPTION
• ON/OFF: to turn the meter on and off.
• ZERO/CFM: this is a bi-functional key. Just press to zero the meter prior to measurement,
or confirm edited values. In
calibration mode
press to confirm factory calibration restore.
• READ/ /TIMER: this is a multi-functional key. In
measurement mode
, press to make a
measurement, or press and hold for three seconds to start a pre-programmed countdown prior
to measurement. In
GLP mode
press to view the next screen.
• CAL CHECK: this is a bi-functional key. Just press to perform the validation of the meter, or
press and hold for three seconds to enter
calibration mode
.
• RANGE/GLP/ : this is a multi-functional key. In
measurement mode
, press to change the
parameter. Press and hold for three seconds to enter
GLP mode
. In
calibration mode
press to
edit the date and time.
OPERATING MODES
•
Measurement mode:
default operation mode, enables both validation and measurement.
•
Calibration mode:
may be entered
by keeping CAL CHECK pressed
for three seconds (
the
“CAL”
tag appears), it enables calibration of the instrument.
•
GLP mode
may be entered by pressing RANGE/GLP/ (“GLP” appears), it enables
consulting of user calibration date or restore factory calibration.
1) RANGE/GLP/ key
2) CAL CHECK key
3) ZERO/CFM key
4) READ/ /TIMER key
5) ON/OFF key
6) Liquid Crystal Display (LCD)
7) Cuvette alignment indicator
8) Cuvette holder
FUNCTIONAL DESCRIPTION
INSTRUMENT DESCRIPTION