beautypg.com

Hanna Instruments HI 96101 User Manual

Page 30

background image

30

DATE INPUT

or

• Place the corresponding CAL CHECK

TM

Standard

Cuvette B into the cuvette holder:

pH: B, HI 96710-11
Free Chlorine: B, HI 96701-11
Total Chlorine: B, HI 96711-11
Cyanuric Acid: B, HI 96722-11
Iodine: B, HI 96718-11
Bromine: B, HI 96716-11
Iron LR: B, HI 96746-11

Ensure that the notch on the cap is positioned
securely into the groove.

• Press READ/ /TIMER and the lamp, cuvette

and detector icons will appear on the display,
depending on the measurement phase.

• After measurement the instrument will show for

three seconds the CAL CHECK

TM

Standard value.

Note:

If the display shows “STD HIGH”, the
standard value was too high. If the display
shows “STD LOW”, the standard value was
too low. Verify that all CAL CHECK

TM

Standards Cuvettes, A and B are free from
fingerprints or dirt and that they are inserted
correctly.

• Then the date of the last calibration (e.g.:

”01.08.2009”

) appears on the display, or

“01.01.2009”

if the factory calibration was

selected before. In both cases the year number
is blinking, ready for date input.

• Press RANGE/GLP/ to edit the desired year

(2009-2099). If the key is kept pressed, the
year number is automatically increased.

• When the correct year has been set, press

ZERO/CFM

or READ/ /TIMER to confirm.

Now the display will show the month blinking.