• Fill a beaker up to the 25 mL mark with
the sample, add the content of one packet
of HI 93722-0 reagent and swirl gently
to mix. This is the reacted sample.
• Fill a second cuvette with 10 mL of the
reacted sample, up to the mark, and
replace the cap.
• Place the second cuvette into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
• Press and hold READ/ /TIMER
for three
seconds.
The display will show the
countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.
• Alternatively, wait for 45 seconds then
just press READ/ /TIMER. In both
cases, the lamp, cuvette and detector
icons will appear on the display,
depending on the measurement phase.
• At the end of measurement, the
instrument directly displays concentration
in mg/L of cyanuric acid on the LCD.
- pH 213 (17 pages)
- pH 211R (13 pages)
- pH 21 (9 pages)
- pH 209V (9 pages)
- pH 209R (13 pages)
- HI 99181 (2 pages)
- HI 99171 (2 pages)
- HI 99163N (2 pages)
- HI 99161N (2 pages)
- HI 99141 (2 pages)
- HI 99131 (2 pages)
- HI 99121 (9 pages)
- HI 99104 (2 pages)
- HI 991003 (2 pages)
- HI 991002 (2 pages)
- HI 991001 (2 pages)
- HI 98402 (9 pages)
- HI 98401 (8 pages)
- HI 98308 (2 pages)
- HI 98240 (21 pages)
- HI 98203 (2 pages)
- HI 98202 (2 pages)
- HI 98201 (2 pages)
- HI 98185 (72 pages)
- HI 98172 (52 pages)
- HI 98160N (48 pages)
- HI 98160 (48 pages)
- HI 98150 (17 pages)
- HI 9815 (2 pages)
- HI 981410 (2 pages)
- HI 981408 (2 pages)
- HI 98121 (2 pages)
- HI 98120 (2 pages)
- HI 98113 (2 pages)
- HI 98110 (2 pages)
- HI 98108 (2 pages)
- HI 98107 (2 pages)
- HI 98106 (2 pages)
- HI 98103 (2 pages)
- HI 96108 (2 pages)
- HI 96107 (2 pages)
- HI 96106 (2 pages)
- HI 96104 (28 pages)
- HI 9321 (36 pages)