TOTAL CHLORINE MEASUREMENT

• Fill the cuvette up to the mark with 10

mL of unreacted sample and replace the
cap.

• Place the cuvette into the holder and en-

sure that the notch on the cap is positioned
securely into the groove.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds, the display will

show “-0.0-”. The meter is now zeroed
and ready for measurement.

• Remove the cuvette and add one packet

of HI 93711-0 reagent. Replace the cap
and shake gently for 20 seconds (or
2 minutes in case of seawater analysis).

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for

three seconds.

The display will show

the countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.