beautypg.com

EXFO FTB-5600 Distributed PMD Analyzer User Manual

Page 73

background image

Managing Results

Distributed PMD Analyzer

67

Section Status

High

Depolarization

(HiDep)

The first or last point in

the section shows a

high intrinsic

polarization level that

causes the

measurement to be

imprecise.

The concerned section

shows a strong level of

intrinsic that causes

the measurement to

be imprecise.

The beating length, the

local variations that

have to be measured,

is a lot shorter than the

size of the pulse.

Therefore the

polarization fluctuation

measured by the unit

vanished through the

pulses spatial

averaging and is not

picked up.
To improve the

situation, you can try

to move the section

cursors outside the

region where the

results are invalid,

increase the spatial

smoothing filter value

or taking a

measurement with a

shorter pulse.

Status

In PMD column

In Status column

Suggestion