beautypg.com

EXFO FTB-5600 Distributed PMD Analyzer User Manual

Page 72

background image

Managing Results

66

FTB-5600

Section Status

Status

In PMD column

In Status column

Suggestion

Valid

The PMD value is

shown

The section is valid

Under Range

(UnRg)

The first or last point in

the section has a PMD

value that is too weak

according to the

residual PMD of the

measurement.

A significant portion of

the cumulated PMD

curve for the section is

to weak according to

the measurement

resolution.

Select a lower scale or

a smaller PMD range in

the acquisition

parameters

Over Range

(OvRg)

The first or last point in

the section has a PMD

value that is too high

according to the

selected PMD range.

A significant portion of

the cumulated PMD

curve for the section is

too high according to

the selected PMD

scale.

Select a higher scale or

broader PMD range in

the acquisition

parameters.

Low SNR

(LoSNR)

The dynamics of the

first or last point in the

section are not

sufficient to measure

PMD correctly for the

section.

The dynamics on the

section is not sufficient

to measure the

cumulated PMD curve

correctly.

Lessen the losses

through the

connectors, increase

the number of

averagings or increase

the pulse size.
For more information,

see Use an Optimized
Number of Averagings

on page 127 and Use
an Appropriate OTDR
Pulse Length
on

page 128.