2. Set the distance range, in kilometers, from the list of available values.

You can also type in the value directly. Reducing the range to the part of
the link that you want to measure will increase measurement speed
and decrease the resulting file size. If you do not know the length of the
link, you can rely on the OTDR trace to estimate it.

3. Press Start.

The acquisition starts. You can see the remaining time in the status bar.

Performing a Standard Acquisition

The acquisitions are more optimized for characterizing the link correctly
rather than for the acquisition time. The default PMD scale for this mode is
20 ps. EXFO recommends setting the accuracy to the standard level and
the sensitivity to the medium level for an acquisition of about 30 minutes.

The accuracy of the measurement changes the number of SOPs and
residual PMD. A higher level of SOPs will yield a more accurate result, but
longer acquisition time.

Accuracy

Number of SOPs

PMD Resolution (fs)

Low

50

1333

Medium

100

148

High

200

50