Setting up and Operating your Distributed PMD Analyzer

38

FTB-5600

Performing an Acquisition

The sensitivity of the measurement allows you to increase the PMD
measurement range. A greater sensitivity increases the number of averages
and will take longer to be achieved.

To improve the dynamic range for your acquisition, consider the following:

³

For less than 13 dB, put the sensitivity to low. This means an acquisition
of about 15 minutes with three resolutions.

³

Setting the sensitivity to medium will improve the dynamic range and
means an acquisition of about 30 minutes with three resolutions.

³

Setting the sensitivity to high will give you the optimal dynamic range
and means an acquisition of about 1 hour with three resolutions.

Note: You can obtain the dynamic range of the trace by subtracting the

backscatter level at the fiber end from the injection level accepted range
value.

Sensitivity

Number of

Averages

Electronic

Sensitivity

Low

500

Low APD gain

Medium

1000

High APD gain

High

2000

High APD gain

Difference between

injection and OTDR

trace level where

the PMD curve ends