Using a Light Source (ELS-500 and EOT-500) or VFL

19

To use the actual source output power as reference (EOT-500 only):
1.
Connect a fiber between the source and power meter of same unit.
2. Activate the source at the desired wavelength.
3. Match the source and power meter wavelengths:

³

If the source emits an auto-wavelength signal
or is in auto-switching mode (see
Automatically Detecting Wavelength on
page 14 and
Using Auto-Switching Mode on
page 17), t
he power meter automatically
matches the source wavelength.
OR

³

Press

to switch between calibrated wavelengths.

4. From normal operating mode, hold down

and

. The unit displays PREF

for a few seconds while saving the value, then returns to normal mode.

To send the source power value:
1.
Activate the source.
2. Change the modulation signal to AUTO or auto-switching.

If you use auto-switching mode, the power value of each wavelength will
automatically be sent.

3. Hold down

and

to send the power value with an auto-wavelength

signal. The unit displays Send for a few seconds, then returns to normal mode.