beautypg.com

0 specifications and troubleshooting tips, 0 accessories parts – ENMET RECON Series User Manual

Page 13

background image

R

ECON

ENMET Corporation

11


6.0 Specifications and Troubleshooting Tips

6.1 Specifications

Electrical Power

Working Voltage

3.6

V

DC

Lithium battery,1200 mAh Rechargeable

Working time continuous, non-alarm conditions

Combustible gas: Approximately 8 hours

Toxic gas: Approximately 30+ hours

Charge time

4 to 6 hours

Operation

Temperature:

–20

°

C to +50

°

C

Relative Humidity

5 to 95% RH, non-condensing

Mechanical

Dimensions:

2.4 x 3.9 x 1.3 inches(60 x 100 x 33 mm)

Weight:

5 oz. (140g) including battery

Ingress Protection

IP 65

Approval

Designed to be intrinsically safe

N

OTE

:

All specifications stated in this manual may change without notice.

6.2 Troubleshooting Tips

M

ALFUNCTION

P

OSSIBLE

R

EASONS

S

OLUTIONS

Unable to Turn on the detector

low voltage

Charge Battery

Detector Fault

Contact ENMET Corporation

Circuit malfunction

Contact ENMET Corporation

No reaction to the tested gas

Exhausted Sensor

Replace Sensor

Circuit malfunction

Contact ENMET Corporation

Inaccurate indication

Exhausted sensor

Replace the sensor

Overdue Calibration

Calibrate

Wrong time display

Exhausted battery

Charge and reset the time

Electromagnetic disturb

Reset the time

Zero calibration function failed

Too much drift

Calibrate or replace the sensor

7.0 Accessories Parts

Recon is supplied with a belt clip(4) and an alligator-type clip(1) with a D-ring(2). These can be attached to the instrument as
shown below.

010101Belt Clip & Alligator

Type Clip Attached

Alligator-Type Clip

2.

D-Ring

1.

Alligator-Type Clip

4.

Belt Clip

Attached

3.

Fastening Screw