5.1.3 Calibration Point Adjustment

In this mode, display is the last concentration of gas used to calibrate the

instrument. Press

UP

or

Down

button to modify the display.

N

OTE

:

Not used with Recon Oxygen units

N

OTE

:

20.9% Oxygen is the only calibration level to be used with Recon Oxygen units

Press the

S

ET

button to accept the level or within 1 minute, the

RECON

will accept the present concentration as the calibration gas concentration.

The display will change to 0..

Open the valve on the regulator

Within 30 seconds, if the

RECON

detects the calibration gas, it will start up the

calibration date analysis procedure by itself. Continue gas flow.

In 2 to 3 minutes the

RECON

will automatically adjust the conversion

arithmetics to complete the gas response process. Then the

RECON

saves the

best arithmetics to complete calibration.

W

ARNING

:

The

RECON

must self calibrate, please avoid touching any

buttons, otherwise the accuracy will be effected badly.

After that it enters into the L-alarm set up. Press

P

OWER

button twice to

advance and exit.

If the

RECON

can not reach half the calibration figure within 30 seconds, or the gas concentration is outside the maximum

drift range of the sensor, the E icon will light and 0000 will be displayed. This means the gas is not proper for calibration or the
sensor is depleted.

Verify calibration gas is correct. If correct retry calibration sequence. If failure occurs sensor may need to be replaced.

Remove calibration cover from instrument.

To exit calibration menu.

Press

S

ET

button 3 times and then the

P

OWER

button 4 times.

Press the

S

ET

button then

Press

P

OWER

button