2 alarm levels adjustment of the recon, 3 battery charging, 4 sensor replacement – ENMET RECON Series User Manual
Page 12: Table1: recon factory set parameters

R
ECON
ENMET Corporation
10
5.2 Alarm Levels Adjustment of the RECON
5.2.1 Set up of L-alarm (Low Alarm)
With the
RECON
Off press and hold the
P
OWER
and
S
ET
buttons until unit activates.
When 0000 is displayed, enter the password, factory set to 1270, by using the
UP
button to increment the digit and the
S
ET
button to accept the indicated digit and move the cursor. When the correct password is entered, the display will advance.
Press the
P
OWER
button to advance to alarm set menu.
In this mode, the L-alarm level can be adjusted. Press
UP
or
Down
button to adjust the flickering figure according to
your
need, and then press
S
ET
button to complete the set up. After that, it enters the H-alarm set up.
N
OTE
:
The display will not allow an unacceptable alarm level.
5.2.2 Set up of H-alarm (High Alarm)
01In this mode, the H-alarm level can be adjusted. Press
UP
or
Down
button to adjust the flickering figure according
to your need, and then press
S
ET
to complete the set up. After that the detector will automatically turn off.
N
OTE
:
The display will not allow an unacceptable alarm level.
5.3 Battery Charging
When the battery voltage is low, the
RECON
will not function properly. To recharge the batteries of the
RECON:
The
RECON
must be Off, connect the AC connector plugs of the charger to a 110
V
AC
or 220
V
AC
outlet and plug the charging
cable into the
RECON
. The detector will activate automatically and indicates that it is being charged. When the charging
cycle is completed, the display will indicate full charge.
W
ARNING
:
D
O
N
OT
start the battery
charge process with the
RECON
power on; otherwise the charge speed will be
adversely affected. Always charge the battery in normal ambient air. Do not charge the
RECON
in a hazardous
area to be monitored, to avoid fire or explosion. Disconnect the charger after 24 hours, failure to do so will
damage the battery.
Typical expected battery life and time to recharge:
RECON
for CO, H2S, O2
30+ Hours, non-alarm condition: Recharge time: approximately 4 – 6 hours
RECON
for %LEL CH4
8 hours, non-alarm condition: Recharge time: approximately 6 hours
N
OTE
:
To maintain proper battery function and life the Recon should be operated once a week for 6 to 8 hours and then placed
on charge.
5.4 Sensor Replacement
Sensor life in the
RECON
is typically: 2 years for Carbon Monoxide & Hydrogen Sulfide and 1year for Oxygen and
Combustible.
The sensor needs to be replaced when the
RECON
can no longer be calibrated or if the sensor becomes erratic.
Contact
ENMET
for sensor replacement information.
Table1: RECON Factory Set Parameters
Model
Gas
Range
Low Alarm High Alarm Low Alarm Range High Alarm Range
TWA
STEL
Cal Gas
R
ECON
/EX Methane
0-100%LEL 10%LEL
50%LEL
10%LEL – 25%LEL
25%LEL – 80%LEL
N/A
N/A
50%LEL
R
ECON
/HS Hydrogen Sulfide
0-100ppm
10ppm
15ppm
5ppm – 15ppm
15ppm – 30ppm
10ppm
15ppm
40ppm
R
ECON
/CO Carbon Monoxide 0-1000ppm 35ppm
100ppm
25ppm – 100ppm
100ppm – 500ppm
25ppm
200ppm 500ppm
R
ECON
/O2 Oxygen
0-30%vol
19.5%vol
23.5%vol
16%vol – 19.5%vol
22.5%vol – 24%vol
N/A
N/A
20.9%