9. General Specifications

Display: 3¾ digit LCD with a maximum reading of 3999.
Range Control: Auto range or manual range control.
Polarity: Automatic negative polarity indication.
Zero Adjustment: Automatic
Overrange Indication: Only the “OL” display.
Low Battery: The symbol “ ” is displayed when the battery
voltage is below approximately 2.4V.
Auto Power Off: 30 minutes after stopping the switch or no key-
input, the meter automatically enters the power off mode.
Safety Standards:

EMC/LVD. The meter is up to the standards

of IEC1010 pollution degree 2; overvoltage category II (double
insulation).
Operating Environment: Temperature 32

O

- 104

O

F (0

O

to 40

O

C),

humidity <80% RH.
Storage Environment: Temperature –4

O

- 140

O

F (–20

O

to 60

O

C),

humidity <85% RH.
Power: Two 1.5V “AA” batteries.
Dimensions: 6 3/32”/155mm (H) x 2 15/16”/75mm (W) x 1 5/16”/
33mm (D)
Weight: Approx. 9.2oz./260g. (including battery and holster)

Detailed Accuracy Specifications

Accuracies are ± (% of reading + number in last digit) at 73

O

F

±41

O

F/23

O

C ±5

O

C, <75% RH.

The detailed accuracy specifications are shown in Tables 7 to 18.

Table 7. DCV Detailed Accuracy Specifications

-33-

+

Range

Resolution

Accuracy

Note

400mV

0.1mV

±(0.5% rdg + 5)

Input Resistance: 10M

Overload Protect: 400mV
range 250V rms. The rest
600VDC or 600VAC rms.

4V

1mV

40V

10mV

400V

100mV

600V

1V

±(0.8% rdg + 5)