Interface parameters

The CVC 3000 E controller is equipped with a serial interface (RS 232C,

nine-pin Sub-D-plug).
+

Plug-in or remove the cable (cable RS 232C) from the interface only if

the equipment is switched off.

+

The interface is

not electrically isolated from the measuring circuit.

+

For optimal electromagnetic compatibility assemble an interface filter

(cat. no.: 638235).

The controller is fully operable via the serial interface. Measuring results,

preselections and the status of the controller can be read at any time.

The factory-set read and write commands are completely compatible

with the VACUUBRAND CVC 2000 controller (see sections ”Read / Write

commands CVC 2000”). An extended instruction set is available using the

command ”CVC 3” (see sections ”Read / Write commands CVC 3000”).

Setting of the interface

Set the interface parameters directly at the CVC 3000 E controller. The

factory set values are underlined.

Edit and confirm the interface parameters in the ”Configuration” menu in

”RS-232” submenu using the selection knob.

Baud: 2400, 4800, 9600 or 19200

Parity: 8-N-1, 7-O-1 or 7-E-1

Handshake: Off, Xon-Xoff or RTS-CTS

Remote: On or Off

Timeout: Sending 1s, receiving 10s.

Readjustment of the controller CVC 3000 E

See section “Readjustment”, pg. 47.

Drain the solvent and dispose of in accordance with regulations. Re-

peat cleaning if necessary.

Rinse the measurement chamber several times with alcohol in order

to remove all solvent residues.

Allow the pressure transducer to dry.

Readjust the pressure transducer if necessary.