Notes on selecting the function – VACUUBRAND CVC 3000 E User Manual
Page 30
The CVC 3000 E controller can be adapted to the specific application by
choosing the appropriate function depending on the connected compo-
nents and the requirements of the application.
Automatic detection of the components
When switching on the controller, the configuration of the connected com-
ponents is checked automatically.
Connected components (e.g., pumps, gauge heads 3000 series, valves)
are
detected automatically and used and supervised until the controller
is switched off. Switch off/on the controller to renew the configuration.
The last mode of operation and the preselected values (e.g., for pres-
sure, speed or time for switching off) are stored.
If the preselections are chosen appropriately, it is possible to start imme-
diately if similar operating conditions are desired.
The controller features
five functions and one configuration menu,
see section ”Menu guide”. Each of these functions involves different menu
options, which are presented automatically and reflect the connected
components.
The types of components connected (e.g., valves) de-
termine the active menu items.
Changing the function:
➨
Switch controller on.
➨
Press ”START/STOP” key to terminate control in case control is run-
ning (e.g., if ”Autostart” is activated).
➨
Press ”MODE” key.
➨
Select function with knob and press to confirm.
+
Depending on the selected function and system components, the con-
troller provides different operating control, as follows:
’’Pump down’’
• Switches a pump or an in-line valve depending on preselected pres-
sure and time settings.
• Coolant valve
”Vac control”
• With pressure preselection, switches a pump or an in-line valve to
maintain that pressure in two-point control.
Notes on selecting the function