Safety during operation, Notice – VACUUBRAND CVC 3000 E User Manual
Page 17

Safety during operation
➨
Adopt suitable measures to prevent the release of
dangerous, toxic, explosive, corrosive, noxious or pol-
luting fluids, vapors and gases.
+
Never operate this controller if it has a damaged cord
or plug. If the controller is not working properly, has
been dropped or has fallen into water, contact your
service provider.
+
Prevent any part of the human body from coming into
contact with vacuum.
+
Attention: At pressures above approximately 795 Torr
(1060 mbar) the pressure reading becomes incorrect
due to saturation of the pressure transducer. The dis-
play flashes. Release pressure immediately!
Risk of
bursting!
• You must take suitable precautions to prevent any dan-
gerous situation from arising if the controller switches
an in-line valve or a coolant valve, switches a vacuum
pump (in combination with a VMS module), or opens a
venting valve.
•
Attention: If the controller is set to Autostart, the pro-
cess will start immediately after a power failure with-
out pressing any further key. It is your responsibility to
ensure that automatic start-up of the system will not
lead to any dangerous condition. Provide appropriate
safety measures. Check prior to starting the process
whether the option ”Autostart” (menu: configuration) is
enabled.
Electronic equipment is never 100% fail-safe. This may
lead to an ill-defined status of the equipment or of other
connected devices. Provide appropriate protective mea-
sures to allow for the possibility of failure and malfunction.
The protective measures must also allow for the require-
ments of the respective application.
NOTICE