Technical data
Technical data of controller
Controller
CVC 3000 E
Pressure transducer
ceramic diaphragm (alumina),
capacitive, absolute pressure,
gas type independent
Display
LCD graphic display, illuminated
Pressure units / scale (selectable)
Torr, mbar or hPa
Measuring range (absolute)
810 - 0.1 Torr (1080 - 0.1 mbar)
Maximum control range (absolute)*
795 - 0.1 Torr (1060 - 0.1 mbar)
Resolution
0.07 Torr (0.1 mbar)
Maximum permissible pressure at pres-
sure transducer (absolute)
1125 Torr (1.5 bar)
Maximum permissible temperature of
gaseous media**
continuous operation: 104°F (40°C),
for short periods (less than 5 minutes)
up to 176°F (80°C)
Measurement uncertainty (absolute)
after careful adjustment and at constant
temperature
<
± 0.75 Torr (1 mbar)
Temperature coefficient
<
± 0.07 mbar/K (0.05 Torr/K)
Ambient temperature range (operation)
50°F to 104°F (10°C to +40°C)
Ambient temperature range (storage)
14°F to 158°F (-10°C to +70°C)
Permissible relative atmospheric mois-
ture during operation (no condensation)
30% to 85%
Maximum permissible range of supply
voltage ( ±25% )
24V DC
Maximum power draw
3.4W (140mA at 24V DC)
Maximum permitted current of connect-
ed valves (connected components)
4A
Degree of protection IEC 529
(front side)
IP 42
Connections to vacuum pump and to
application
connection for PTFE tube
I.D. 3/8”, O.D. 5/16” (10 / 8 mm)
Venting connection
hose nozzle for hose I.D. 3/16” (4-5 mm)
Maximum admissible pressure at vent-
ing connection
17.4 psi (1.2 bar) absolute