Program function – VACUUBRAND CVC 3000 E User Manual
Page 38
Program function
➨
Permits ten programs to be defined and stored, each with up to
ten program steps with preset values for vacuum and time.
+
Edit:
Use to define the preset values for the process run:
Time: Defines either the process runtime for each program step to
reach a preset vacuum level or, if programming a ”Step”, the runtime
after having achieved the vacuum level. The total process runtime is
shown in the base line. Attention: A preset runtime of 99:59:59 hours
in the final program step will cause the process to run endlessly. Ter-
minate the process by pressing the ”STOP” key.
Vacuum: Vacuum level to be attained.
Venting valve: Operating a venting valve to reach a preset vacuum
level.
”Step”: ”Step” causes pump down as fast as possible to the preset
vacuum level. As soon as the vacuum level is reached the time meter
starts running.
+
Open: Choose program (Programs 0 - 9).
+
Store: This command stores an edited program or the program of the
last process to one of the storage spaces 0 - 9.
+
Hysteresis: The ”Hysteresis” is the control bandwidth of the two-point
control. A too small hysteresis will lead to frequent switching of the
valve or the pump. A too large hysteresis will lead to imprecise con-
trol. Suggested ”Hysteresis” values are stored in the controller (setup
”Auto”) and are adapted automatically to the preset pressure. The hys-
teresis can be adapted at any time.
The ”Hysteresis” is adjustable in a range of 1-225 Torr (1-300 mbar) or
can be set to ”Auto”.
+
Delay: ”Delay” determines the time the coolant valve remains open af-
ter the process has been stopped. Determines also the time the pump
(only with VMS module and in-line valve) remains running after the
process has been stopped.
The ”Delay” is adjustable in a range of 1-300 minutes or can be set to