Troubleshooting – VACUUBRAND MD 1C VARIO-SP User Manual
Page 56
page 56 of 87
Fault
Possible cause
Remedy
❑ Pump does not
start or stops im-
mediately.
➨ Electrical power cord
not plugged in, electri-
cal supply failure?
✔ Plug in power cord.
Check fuse. Check elec-
trical supply voltage.
➨ Device fuse blown (if
applicable)?
✔ Identify cause of failure.
Replace device fuse.
➨ Overpressure in outlet
line?
✔ Remove blockage in line,
open valve, or reduce
overpressure.
➨ Motor overloaded?
✔ Allow motor to cool down,
identify and eliminate
cause of failure. Manual
reset is necessary. Switch
off pump or unplug.
❑ Pump does not
achieve its ultimate
vacuum or usual
pumping speed.
➨ Leak in the pipeline or
vacuum system?
✔ Check pump directly -
connect vacuum gauge
directly at pump inlet -
then check connection,
pipeline and vacuum
system if necessary.
➨ Long, narrow vacuum
line?
✔ Use lines with larger di-
ameter, length as short as
possible.
➨ Pump has been ex-
posed to condensate?
✔ Allow pump to run for
some minutes with atmo-
spheric pressure at the
inlet to purge.
➨ Deposits have been
formed inside the
pump?
✔ Clean and inspect the
pump heads.
➨ Diaphragms or valves
damaged?
✔ Replace diaphragms and/
or valves.
➨ Outgassing substances
or vapor generated in
the process?
✔ Check process parame-
ters.
➨ Only VARIO-SP ver-
sion: Pump tempera-
ture too high (motor
speed reduced)?
✔ Ensure sufficient cooling
of the pump or reduce
inlet pressure.
Troubleshooting