In case of condensation, Notice – VACUUBRAND MD 1C VARIO-SP User Manual
Page 52
page 52 of 87
In case of condensation
In case of
condensation: Check the liquid lev-
el in the catchpots during operation. Check the
liquid level in the catchpots regularly. Do not
allow the catchpots to overfill. Drain catchpots
in time to avoid overflow. Install a level sensor
(see „Accessories“, pg. 54) for monitoring,
if necessary (VACUUBRAND controller CVC
3000 or VNC 2 is required).
The maximum liquid level is at approximately
80% of the total filling level to avoid problems
when removing the catchpots.
Drain catchpots.
+
Important: Comply with regulations when
disposing of solvents/condensates. Recy-
cle if possible; purify if contaminated.
Removing the catchpots:
Catchpot at outlet:
Remove joint clip. Remove catchpot and drain
condensate.
Catchpot at inlet:
Admit air or inert gas (via the pump inlet) to
restore atmospheric pressure in the catchpot
before attempting removal. Remove joint clip.
Remove catchpot and drain condensate.
Reattach drained catchpots.
NOTICE
✘
✔
80%
max.