• Only install the optional coolant valve in the
supply line of the exhaust waste vapor con-
denser.
•
Note: Install the hoses of the cooling system
in a way to avoid the flow / dripping of con-
densed water onto the pumping unit (espe-
cially cables and electronic parts, see also
IP degree of protection, „Technical data“,
• Ensure that the
coolant outlet tubing is always unob-
structed and that it cannot get blocked.
• Maximum permissible coolant pressure at the exhaust
waste vapor condenser: 87 psi (6 bar) absolute. Outlet
flow must always be unhindered.
• Comply with the maximum permissible coolant pres-
sures of additional components in the coolant circuit
(e.g., coolant valve).
• Avoid overpressure in the coolant circuit (e.g., caused
by blocked or squeezed coolant hoses).
✘
✔
EX
IN
Permissible range of coolant temperature at the exhaust
waste vapor condenser: 5°F to 68°F (-15°C to +20°C)
Check hose connections prior to starting operation of the
cooling system.
Check coolant hoses regularly during operation.
NOTICE
MD 1C (AC motor)
Plug in the power cord.
• Check the power source and the pump’s
rating plate to be sure that the power source
and the equipment match in voltage, phase,
and frequency.
Switch the pump on.