The supply cable (pumps with AC motor) may be fitted
with a molded European IEC plug or a plug suitable for
your local electrical supply. The cable contains wires
color coded as follows: green or green and yellow:
ground; blue or white: neutral; brown or black: hot.
+
Due to the high compression ratio, the pump may gen-
erate overpressure at the outlet. Check pressure com-
patibility with system components (e.g., exhaust pipe-
line or exhaust valve) at the outlet.
+
Do not permit any
uncontrolled pressurizing. Make
sure that the exhaust pipeline cannot become blocked.
If there is an exhaust isolation valve, make sure that you
cannot operate the equipment with the valve closed to
avoid a
risk of bursting!
+
Keep the electrical power cord away from heated sur-
faces.
• Provide a firm, level platform for the equipment. Check
that the system which you are going to evacuate is
mechanically stable. Check that all fittings are secure.
Ensure a stable position of the pump without any me-
chanical contact other than the pump feet.
• Comply with
maximum permissible pressures at in-
let and outlet and with
maximum permissible pres-
sure differences between inlet and outlet. See section
„Technical data“, pg. 25. Do not operate the pump
with overpressure at the inlet.
• Avoid overpressure of more than 17.5 psi absolute (1.2
bar absolute) in the event that gas or inert gas is con-
nected to the pump, to the gas ballast or to a venting
valve.
•
Note: Flexible elements will shrink when evacuated.
• Connect hoses gas tight at inlet and outlet of the pump.
• Ensure that no foreign objects can be drawn into the
pump.