Notice – VACUUBRAND MD 1C VARIO-SP User Manual
Page 40
page 40 of 87
Catchpot:
The catchpot is coated with a protective layer
to protect against shattering in case of break-
age or implosion.
➨
Assemble the catchpot at the exhaust waste
vapor condenser using a joint clip.
over-
pressure
safety
relief
device
+
Check the overpressure safety relief device
at the exhaust waste vapor condenser regu-
larly; replace if necessary. Check especially
for deterioration, coalescence and cracks.
➨
Assemble the hose nozzles for coolant in-
let and coolant outlet tubing at the exhaust
waste vapor condenser (hose nozzles for
tubing I.D. 1/4”-5/16” (6-8 mm)).
➨
Attach the tubing of the coolant circuit to the
respective hose nozzles at the waste vapor
condenser.
➨
Prevent the discharge of dangerous gases and vapors
to the surrounding atmosphere. If appropriate, connect
the exhaust line to a suitable treatment system.
Check all hose connections prior to starting
operation of the cooling system.
Secure coolant hoses at the hose nozzles
(e.g., with hose clamps) to prevent their acci-
dentally slipping off.
EX
IN
+
Never block the gas outlet (hose nozzle for
tubing I.D. 3/8” (10 mm)). The exhaust hose
has always to be unobstructed and with-
out back pressure to enable an unhindered
discharge of gases and protect the pump
valves from damage.
NOTICE
outlet