Separator (ak) at the inlet, Connecting the outlet, Notice separator (ak) at the inlet – VACUUBRAND MD 1C VARIO-SP User Manual
Page 38: Connecting the outlet notice
page 38 of 87
• A power failure may cause accidental ventilation of the
pump, especially if the gas ballast valve is open. If this
constitutes a potential source of danger, take appropri-
ate safety measures.
When assembling, ensure
vacuum-tightness. After as-
sembly, check the whole system for leaks.
Use a suitable valve (see „Accessories“, pg. 54) to iso-
late the pump from the vacuum application. This is to allow
the pump to warm up before pumping condensable vapors
and to clean the pump after use before it is switched off.
NOTICE
Separator (AK) at the inlet
The separator at the inlet protects against
droplets and particles from entering the pump.
+
Enhances lifetimes of diaphragms and
valves.
+
Improves vacuum performance in applica-
tions with condensable vapors.
Catchpot:
The catchpot is coated with a protective layer to protect
against shattering in case of breakage or implosion.
➨
Assemble the catchpot at the inlet using a joint clip.
Outlet via hose nozzle for tubing I.D. 5/16”
(hose nozzle DN 8 mm) or via hose nozzle for
tubing I.D. 3/8” (hose nozzle DN 10 mm) at ex-
haust waste vapor condenser EK.
Connecting the outlet
NOTICE
outlet
EK
➨
If dangerous or polluting fluids could be released at
the outlet, install an appropriate system to catch and
dispose of those fluids.