Replacing diaphragms and valves, See section “re, Placing diaphragms and valves”, pg. 91) – VACUUBRAND MV 10 NT VARIO User Manual
Page 91
page 91 of 126
+
Please read section ”Replacing diaphragms and valves” com-
pletely before starting maintenance.
The pictures may show other versions of pumps. This does not change
the method of replacing diaphragms and valves.
➨
Never operate the pump if covers or other parts of
the pump are disassembled.
➨
Before starting maintenance,
disconnect the electri-
cal power cord. Wait two minutes after isolating the
equipment from AC power to allow the capacitors to
discharge.
➨
Ensure that the pump cannot be operated accidentally.
➨
Note: The pump might be contaminated with the pro-
cess chemicals that have been pumped during opera-
tion. Ensure that the pump is decontaminated before
maintenance.
➨
Avoid the release of pollutants.
+
Never operate a defective or damaged pump.
+
Take adequate precautions to protect people
from the effects of dangerous substances
that may have contaminated the pump and
may be released upon disassembly. Ensure
that the maintenance technician is famil-
iar with the safety procedures which relate
to the products processed by the pumping
system.
Use appropriate protective clothing, safety
goggles and protective gloves.
Replacing diaphragms and valves