Electrical connection, Notice electrical connection – VACUUBRAND MV 10 NT VARIO User Manual
Page 37
page 37 of 126
+
Maximum permissible pressure at the pres-
sure transducer: 21.8 psi (1.5 bar) absolute.
+
Attention: At pressures above approxi-
mately 795 Torr (1060 mbar) the pressure
reading becomes incorrect due to satura-
tion of the pressure transducer. The display
will flash. Release pressure immediately!
Risk of bursting!
maximum
21.8 psi
(1.5 bar)
absolute
✘
Condensate and deposits will affect the mea-
surement results.
Position the pressure transducer and its vacu-
um line in such a way that condensate cannot
flow towards the pressure transducer. Clean
the pressure transducer, if necessary. See
section “Cleaning the pressure transducer”,
Connect the CVC 3000 controller to the
VACUU•BUS cable of the NT VARIO pump.
Plug in the power cord.
• Check the power source and the pump’s
rating plate to be sure that the power source
and the equipment match in voltage, phase,
and frequency.
Switch the pump on.
O
➨ I
NOTICE
Electrical connection