Shutdown & storage, Notice – VACUUBRAND MV 10 NT VARIO User Manual
Page 41
![background image](/manuals/715235/41/background.png)
page 41 of 126
Shutdown & storage
The pump can be switched off under vacuum.
Short-term:
- Has the pump been exposed to conden-
sate?
Allow the pump to continue to run at atmo-
spheric pressure for a few minutes.
- Has the pump been exposed to media which
may damage the pump materials or form
deposits?
Check and clean pump heads if necessary.
- Has the pressure transducer been exposed
to media which may form deposits?
Clean pressure transducer if necessary.
Long-term:
- Take measures as described above regard-
ing short-term shutdown.
- Separate the pump from the application.
- Close inlet and outlet ports (e.g., with trans-
port caps).
- Store the pump under dry conditions.
NOTICE
NOTICE
Condensate?
O
➨ I
(5-10min)
atm