beautypg.com

Application examples, Vacuum for filtration and suction, Assembly of a vacuum system – VACUUBRAND MV 10 NT VARIO User Manual

Page 61

background image

page 61 of 126

Assembly of a vacuum system

+

Assemble vacuum connection lines between vacuum pump, CVC

3000 and vacuum application.

+

Assemble the vacuum connection line between the controller and the

vacuum application, if the controller is not part of a pumping unit.

+

Assemble electrical connections.

+

Connect coolant if necessary.

Vacuum for filtration and suction

+

Select ”

Pump down” function.

+

If necessary, set value for ”Speed” (high or low pumping speed). Re-

duce the speed in case of fast sedimentation (formation of a filter cake).

+

Set ”Set vacuum” to a value which provides adequate suction but does

not lead to evaporation of the solvent.

+

Start process by pressing ”START/STOP” key.

alternatively:

+

Select ”

Vac control” function.

+

Set ”Set vacuum” to a value which does not lead to evaporation of the

solvent.

+

If necessary, set value for ”Speed” (high or low pumping speed). Re-

duce the speed in case of fast sedimentation (formation of a filter cake).

+

Set ”Maximum” pressure setting so that pump will switch off at the end

of the filtration process, or in the event that a filter cake cracks, leading

to a sudden pressure increase.

+

Start process by pressing ”START/STOP” key.

+

Tip for filtration: Adjust preset pressure to a value well above the

boiling pressure of the solvent (e.g., for water >>15 Torr (20 mbar)).

Set the maximum pressure to e.g., 375 Torr (500 mbar). Once the fil-

tration has finished, the pressure increases and the pump is switched

off automatically.

Application examples