General view basic setup vacuulan, General view basic setup vacuu•lan – VACUUBRAND PC 620 User Manual
Page 29
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 999029 / 01/10/2009
page 29 of 48
General view basic setup VACUU•LAN
connect coolant
valve
setting process parameters
➨ pressure value (condition for automatic shut down)
➨ upper set point (condition for restart)
➨ time for automatic shut down
setting the basic setup
basic setup VACUU•LAN
yes
no
switching coolant
valve?
start: pump down
☞
☞
☞
☞
☞ The pump (and the coolant valve if applicable) is switched
off if the pressure is below the lower pressure value
longer than the time for automatic shut down.
☞
☞
☞
☞
☞ The pump starts automatically if the pressure is higher
than the upper set point.