Setting the lower pressure value (condition for automatic shut
down):
➨ Press p▲ or p▼.
☞ ”Set p” and the lower pressure value are displayed for approx.
1s.
☞ With a second tip within one second or continuous press-
ing:
➨ Set the lower pressure value by pressing key p▲ or p▼.
☞ The lower pressure value is stored as the new set point
when the key is released.
Note: It is recommended to select a pressure value which is
approx. 10 mbar higher than the ultimate pressure value achiev-
able in the system.
Setting the process parameters
Setting the time for automatic shut down:
➨ Press key ▲ or ▼.
☞ ”Set”, the clock symbol and the time for automatic shut down
are displayed for approx. 1 s.
☞ With a second tip within one second or continuous press-
ing:
➨ Press key ▲ or ▼ to set the value for the time for shut down
(1 - 200 min).
☞ The value for the shut down time is stored as the new set
point when the key is released.
☞ If the actual pressure falls below the lower set point, the
time for automatic shut down starts to run. The time for au-
tomatic shut down is reset if the actual pressure rises higher
than the lower set point (with a hysteresis of 3 mbar).
The process control stops when the time for shut down has
elapsed. The vacuum pump is switched off and the coolant valve
is closed.
☞ Pump starts again in case of great pressure increase.
☞ If the actual pressure is rising above the upper set point, the
pump starts again (corresponds to starting by pressing key
START/STOP).
☞ The process control can be stopped manually at any time
by pressing key START/STOP.
Setting the upper pressure value (condition for restart):
➨ Keep key MODE pressed and press key ▲ additionally.
☞ ”Set p”, ”arrow up” and the upper pressure value are dis-
played for approx. one second.
247
mbar
67
mbar
MODE
67
mbar
p
015
☞ With a second tip within one second or continuous press-
ing:
➨ Set the upper pressure value by pressing key ▲ or ▼.
☞ The upper pressure value is stored as the new set point
when the key is released.